Name
Plan

Method
The independent calculation is performed by a physicist or dosimetrist using the standard TMR based formalism using a reference calculation point from the treatment planning system for each beam. Values of point dose, energy, feld size, effective depth, SAD and wedge angle are entered from the RTP system. The program looks up values for the TMR, Sc, Sp, WF, and OAR to calculate the MU for dose at the calculation point. It is given by: MU = D/(TMR x Sc x Sp x OAR x WF x inv2). Acceptance criteria for plan vs calculation differences follows AAPM guidelines for MU calculations as outlined in Task Group Report 71 (2014). Disclaimer: The responsibility of ensuring patient safety and accuracy for using this program rests entirely with the user. (Copyright © 2018: medical-physics-tools.org)

beam
linac
energy
wedge
X1
X2
Y1
Y2
block%
depth
SAD
offsetx
offsety
Θw
Dose
weight
MUplan
TMR
OAR
wdgF
Sc
Sp
inv2
MUhand
Δ